Import Protocols - Numaris X
This Job Aid provides the workflow steps to Import Protocols on Numaris X software.
Import Protocols – Numaris X Job Aid Import Protocols – Numaris X Important Recommendations It is recommended to import protocols only when there is allocated time on the MAGNETOMR MR system and no other tasks (scanning, applications) are running. Note: The MR System will be locked while the protocol import function is running and scanning will be prohibited. Notes Import Protocols and Exchange – Entire Protocol Tree, Region, Exam, or Program o Protocols can be imported “From Existing File” or “From Teamplay” MR MAGNETOM System Importing from – must have the same or an older software ver- sion and configuration o Note: older MR MAGNETOM System configurations may need time to automatically convert the protocols being imported. Once the import is completed the conver- sion is indicated by an underlined sequence 1. Import Protocols fom an Existing File a. Insert “USB”, “External Hard Drive” into MR System i. Note: ensure USB, External Hard Drive is from a trustworthy source (or) has been scanned for viruses prior to inserting into the MR system b. Open Dot Cockpit (using one of the options) i. Home Screen Icon CH System Check Dot Cockpit Examination Lock ii. Dot Cockpit Icon – bottom of Measurement Queue SAR Stimu Exam First Vorma Made SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: 05/26/2020 | HOOD05162003089549 Job Aid | Import Protocols - Numaris X Effective Date: 05/26/2020 HOOD05162003089549 c. Select Explorer > Import i. Data Import – select an option to Import Protocols 1. From Existing File 2. From Teamplay ii. Select Area to Import Protocol data 1. New Tree, Existing Tree (Region, Exam, Program) iii. Click Import Icon Dol Cockpit Dot Cockpit administrator mode - use for aarrestation of password protection of trees only Quit Explorer Program Editor Browse Import Export Local Database USER Q WorkflowSimulation > Hip_Dot_Engine_Bilateral › C-Spine Dat Engine > Knoc Dot Engine > Abdomen Dot Engine > Heart >Brain Data Import Whole Body Dat Engine Knee Dot Engine 20 From Existing File Knee Dat Engine From teamplay A Import Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 2 of 5 Job Aid | Import Protocols - Numaris X Effective Date: 05/26/2020 HOOD05162003089549 d. Dialogue Window Opens i. Select “USB”, “External Hard Drive” plugged into MR System ii. Locate and select “ .exar1” file iii. Click “Open” – selection confirmed and Dialogue window closes Dot Cockpit Dot Cockpit administrator mode - use for administration of password protection of trees only Quit Explorer Program Editor Browse mport Export Local Database USE | Open x a << Knee_Dot_Engine_2D > exar.file Search exar.file Organize New folder Work!! Searches Name Date modified Type Hip Videos ] general 20.exart 6/6/2019 3:19 PM EXAR1 File This PC Desktop > Knee Downloads Abdo KA > Hear CustomerSeq > Brain DicomExport ata Import DICOMTestin Whol dotencine Knee From Existing File Knee Dot_En Knee_Dot_E From teamplay excar.file Images Import ScreenCaptur v ‹ File name: general 20.exart Cede"ear1) Open Cancel Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 3 of 5 Job Aid | Import Protocols - Numaris X Effective Date: 05/26/2020 HOOD05162003089549 e. Dot Cockpit Window reopens i. Select Protocol to Import ii. Start Import by selecting Arrow (Shuttle Button) or “Drag and Drop” into Dot Cockpit (User Tree, Region, Exam, or Program) file area. iii. Import Complete A Dot Cockpit Dot Cockpit administrator mode - use for administration of password protection of trees only Quit Explorer Program Editor Browse Import Export Local Database ... 2D.exar1 X general 2D USER Vida-XT-128 Export Q Q WorkflowSimulation knee > Hip_Dot_Engine_Bilateral Knee Dot Engine › C-Spine Dot Engine general 2D > Knee Dot Engine Abdomen Dot Engine > Heart Brain › Whole Body Dot Engine Knee Dot Engine 2D Knee Dot Engine « Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 4 of 5 Job Aid | Import Protocols - Numaris X Effective Date: 05/26/2020 HOOD05162003089549 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 5 of 5
- Import Protocols
- Numaris X