PEPconnect

Sysmex UF-1000i™ Cleaning the Sample Filter Video

This video provides instructions on how to properly clean the sample filter on the Sysmex UF-1000i™ Urine Cell Analyzer.

The sample filter should be cleaned when it's clogged and sample can no longer be aspirated. Start by turning the instrument off. Wait one minute before continuing to allow the pressure to drop. Then open the front cover of the instrument. Locate the sample filter and remove the filter from the adapter. Remove the filter fixture from the upper and lower tubing by unscrewing the upper and lower fittings. Unscrew the filter fixture external thread from the internal thread. And then remove the filter from the filter fixture. Rinse the filter with D ionized water. Check the filter hole for clogs. If clogs are present, rinse the filter again and if clogs cannot be removed by rinsing, it may be necessary to replace the sample filter. Insert the cleaned filter into the filter fixture. Tighten the filter fixture external thread to the internal thread. Attach the filter fixture to the upper and lower tubes supporting braces. And then return the filter fixture to the sample filter adapter. Close the front cover. And turn the instrument on. After replacement process QC to verify accuracy of aspiration, absence of leaks and system performance.

u F-loooi NO 7000 i Cleaning the Sample Filter ( åaning the

  • uf-1000i as needed maintenance
  • uf1000i as needed maintenance
  • clean sample filter
  • replace sample filter